User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: 8522A MIL-STD-45662A Before Data 90 Day Spec DATE: 14-Mar-91 AUTHOR: User Contributed REVISION: 2.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 40 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W #Source: Calibration Manual Dated:11/81 Section 5 #The 24 Hour Spec of the 5700A is used in the TUR computations 1.002 DISP EQUIPMENT SETUP 1.003 DISP WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 DISP Connect the UUT to an AC power source. 1.004 DISP Turn UUT on. 1.004 DISP Warm-up time: 1 Hour 1.004 DISP Ambient temperture: 22C - 24C 1.004 DISP 1.004 DISP Connect the UUT to the UUT IEEE-488 bus. 1.005 DISP Connect the 8522A input high to the 5700A output high 1.005 DISP Connect the 8522A input low to the 5700A output low 1.005 DISP Connect the 8522A source high to the 5700A sense high 1.005 DISP Connect the 8522A source low to the 5700A sense low 1.005 DISP Connect the 8522A guard to the 5700A guard 1.006 DISP DEPRESS THE EXTERNAL GUARD BUTTON ON THE 8522A 1.006 DISP RELEASE THE FRONT/REAR BUTTON ON THE 8522A 1.007 CALL 8522A MIL-STD-45662A 90 Day Spec 1.008 JMP 2.001 1.009 EVAL 2.001 CALL 8522A RESISTANCE 90 Day Spec 2.002 JMP 3.001 2.003 EVAL 3.001 CALL 8522A AC VOLTS 90 Day Spec 3.002 END